Technical Service
Quick Set-up Guide
CAUTION: SEE OPERATIONS MANUAL FOR COMPLETE
INSTRUCTIONS, INCLUDING ALL WARNINGS AND PRECAUTIONS PRIOR TO SETUP.
Initial
setup
- Connect power cord to INVOS Cerebral Oximeter
monitor and power source.
- Make sure back panel master switch is turned
on.
- Connect Preamplifier to INVOS monitor.
Connect Bilateral Reusable SomaSensor Cable to Preamplifier.
For
each patient
- Turn power on by pressing the green button
at the lower left front corner of monitor.
- INVOS System completes a self-test within
30 seconds.
- Degrease forehead with the enclosed alcohol/acetone
wipe to ensure good contact. After forehead is dry, place SomaSensors
bilaterally on patient's forehead. Sensor cable can be connected
to sensors before or after placement. Hold blue connectors, not
the sensor black flex circuit when connecting sensors to cable.
- Press NEW PATIENT or PREVIOUS PATIENT.
- Observe right and left channel measurements.
Set alarm limits, if desired. Observe status messages on screen
for troubleshooting assistance.
- After trending
begins and patient condition is stable (i.e., surgical patient,
prior to induction), set the awake baseline. Press BASELINE MENU.
Press SET BASELINE. The BASELINE and % CHANGE FROM BASELINE values
will be displayed below the real-time value, and you will be automatically
returned to the MAIN MENU.
To download a PDF of this guide, click
here. (151 KB)